The largest database of trusted experimental protocols

Lambda 25 uv vis spectrometer

Manufactured by PerkinElmer
Sourced in United States, Singapore, Italy, United Kingdom

The Lambda 25 UV/VIS Spectrometer is a compact, high-performance double-beam spectrophotometer designed for routine UV/Vis analysis. It features a wavelength range of 190 to 1100 nm and a spectral bandwidth of 1.0 nm. The instrument is equipped with a deuterium and tungsten-halogen light source and includes a high-sensitivity photodiode detector.

Automatically generated - may contain errors

122 protocols using lambda 25 uv vis spectrometer

1

Characterization of Ag-Nanostructure Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
The morphological shape and formation mechanism of Ag-nanostructure materials were characterized by using Transmission Electron Microscope (TEM, JEM-2100, Akishima, Tokyo, Japan) and High-resolution Transmission Electron Microscope (HR-TEM, FEI Talos F200C at 200 kV, Tokyo, Japan) and scanning electron microscopy (SEM, JSM-6700F, Akishima, Tokyo, Japan). The XRD analysis was performed with a Bruker D8 Advance X-ray diffractometer equipped with a Cu Ka radiation source (Karlsruhe, Germany). The measurements of optical properties were conducted by using a PerkinElmer UV/VIS Spectrometer Lambda25, manufactured by PerkinElmer, Ayer Rajah Crescent, Singapore Pte Ltd., and UV-2450 UV-vis Shimadzu spectrophotometer, Kyoto, Japan. Meanwhile, the hydrodynamic size and zeta potential were also measured by dynamic light scatting (DLS) using a Nano-ZS ZEN3600 Malvern Instruments, Worcestershire, UK. Also, the measurements of the cell viability were carried out by using Enzyme-Linked Immunosorbent Assay (ELISA, Thermo Fisher Scientific, Waltham, MA, USA) at 570 nm wavelength.
+ Open protocol
+ Expand
2

Spectrophotometric Analysis of Glutaraldehyde, HEMA, and Chlorhexidine

Check if the same lab product or an alternative is used in the 5 most similar protocols
To prepare 25ml of a solution containing glutaraldehyde 5% and HEMA 35% and chlorhexidine 2%, the amount of 7.7ml of standard solution
of HEMA with purity of 97% and 3.9ml of purified glutaraldehyde 25% with 2ml of 2% chlorhexidine were mixed. To investigate the chemical
reaction of these materials, each solution was first separately detected in an ultraviolet–visible spectrophotometer (UV/VIS Spectrometer lambda 25,
PerkinElmer, USA), and the wavelength-absorption diagram of each substance alone was recorded. Then the mixture of three substances was placed
in the device and a mixture diagram of these substances was recorded. It should be noted that all stages of construction of new material were carried out by an experienced pharmacist.
+ Open protocol
+ Expand
3

Analyzing Film Barrier Properties

Check if the same lab product or an alternative is used in the 5 most similar protocols
The ultraviolet and visible light barrier properties of the films were analysed in a spectrophotometer Perkin Elmer UV/VIS spectrometer Lambda 25 (Waltham, Massachusetts) using transmittance mode in UV-VIS regions (200–800 nm). This analysis was conducted in duplicate. Transparency and UV-barrier property were determined according to the Equations (4) and (5) proposed by Han [32 ]: Transparency=log%T660x
where %T660 is the percent transmittance at 660 nm and x is the film thickness (mm).
UVbarrier=100%T280%T660×100
where %T280 is the transmittance at 280 nm and %T660 is the transmittance at 660 nm.
+ Open protocol
+ Expand
4

Copper(II) Complex Characterization

Check if the same lab product or an alternative is used in the 5 most similar protocols
Copper(II) chloride dihydrate; 3,4-difluorophenylacetic acid; 2-methylpyridine; 3-methylpyridine; 2,2-diphenyl-1-picrylhydrazyl radical (DPPH); ascorbic acid; iron(II) sulfate; hydrogen peroxide; nitric acid; sodium salt of the salmon sperm DNA (SS-DNA); methanol; and DMSO were purchased from Sigma-Aldrich, St. Louis, MO, USA. These chemicals were used without any further treatment. Distilled water was used for the experimental work. The melting points of the synthesized complexes were recorded in a capillary tube using a digital electro-thermal melting point apparatus. Elemental analysis was performed on a Leco CHNS 932. A Perkin Elmer atomic absorption spectrometer A analyst 700 was used to determine the percentage of copper. The electronic absorption spectra (200–800 nm) of the complexes were recorded using a Perkin Elmer UV/Vis spectrometer Lambda 25 in DMSO solvent. A nicolet-6700 FT-IR spectrophotometer (Thermo Scientific, Waltham, MA, USA) was used to record FT-IR spectra in the range of 4000–400 cm−1, adopting the attenuated total reflectance (ATR) technique.
+ Open protocol
+ Expand
5

Hollow Fiber Membrane Filtration Performance

Check if the same lab product or an alternative is used in the 5 most similar protocols
Pure water flux of the hollow fiber membrane was evaluated at a constant temperature of 25 °C using crossflow filtration system in three modules each containing five fibers with one end sealed. Prior to the test, the hollow fiber membranes were soaked in tap water for 30 min and followed by a pre-compression at 1 bar pressure for another 30 min to attain stable pure water volume flux. The pure water flux, Jw was determined using Equation (1) [43 (link)]:
where Jw signifies the permeation flux in L/m2·h. Q represents the amount of permeated pure water in L. A corresponds to the effective area of the membrane in m2 and t is the time required to get the amount of Q in hour. The pure water permeability (PWP) was calculated by using Equation (2) [14 (link)]. TMP is the transmembrane pressure in bar.

The membrane oil rejection percentage (R) was determined based on Equation (3) [44 (link)]. The oil concentrations in the permeate and feed were determined using PerkinElmer UV-Vis Spectrometer Lambda 25 (Waltham, MA, USA) at the maximum absorbance of oily wastewater measured at 223 nm. CP and CF signify the permeate and the feed concentration in ppm, respectively.

+ Open protocol
+ Expand
6

Naringin Deposition Analysis using SSGI

Check if the same lab product or an alternative is used in the 5 most similar protocols
The in vitro deposition of micronized naringin was evaluated using a glass impinger at a single stage (SSGI, European Pharmacopoeia 6.0 instrument, Copley Ltd. Scientific, Mothingam, UK).
A water/ethanol 9/1 v/v mixture was introduced in the upper (7 mL) and lower (30 mL) stages of the SSGI. Hard gelatin capsules (size 2) were filled with 20.0 ± 0.5 mg of naringin powders and introduced into the Turbospin. The vacuum pump was operated at a flow rate of 60 L/min for 5 s. Each deposition experiment was carried out on 10 capsules and repeated in triplicate. A UV spectrometer (UV/Vis spectrometer Lambda 25, Perkin-Elmer instruments, Waltham, MA, USA) was used at a wavelength of 283 nm to quantify the naringin deposited into the upper and the lower stages of the impinger. The analytic method was validated using standard solutions of naringin in the range of 70–400 mg/L. The emitted dose (ED) was gravimetrically determined and expressed as the percentage of the powder at the exit of the apparatus vs. the amount of powder introduced into the capsule. The ratio between the naringin recovered from the SSGI and the naringin emitted from the device was the recovered dose. The fine particle fraction (FPF) was defined as the percentage of the ratio of the naringin recovered from the lower stage of SSGI vs. the total loaded dose.
+ Open protocol
+ Expand
7

Multimodal Characterization of Novel Materials

Check if the same lab product or an alternative is used in the 5 most similar protocols
1H NMR was recorded on a 400 MHz Bruker spectrometer running and analysed using TopSpin software. Deuterated acetone was employed as the reference. UV–Vis was performed on a PerkinElmer UV/VIS Spectrometer Lambda 25. XRD patterns were measured on a PANalytical X'Pert Pro MRD diffractometer using Ni filtered Cu Kα radiation at 40 kV and 40 mA. SEM-EDX measurements were carried out on a JEOL 6400 scanning electron microscope operated at 20 kV. IGA measurements were conducted on a Mettler Toledo TGA spectrometer. For ToF-SIMS samples were fabricated onto clean glass substrates. The samples were then soaked under dry flux in the dark for an hour before the ToF-SIMS measurements were recorded. Data were obtained using an IONTOF ToF.SIMS-Qtac LEIS spectrometer employing an Argon sputter gun for oxygen ion detection.
+ Open protocol
+ Expand
8

UV-VIS Spectroscopy of Volvox MCP

Check if the same lab product or an alternative is used in the 5 most similar protocols
For UV VIS spectroscopy, recombinant V. carteri MCP was adjusted to 600 µM in purification buffer (100 mM Tris-HCl pH 8.0, 150 mM NaCl, 1 mM EDTA, 5% (v/v) glycerol). Spectra were recorded using a PerkinElmer UV/VIS Spectrometer Lambda 25 between 380 and 620 nm.
+ Open protocol
+ Expand
9

Ribera del Duero Cabernet Sauvignon Wine Aging

Check if the same lab product or an alternative is used in the 5 most similar protocols
A red wine, from Ribera del Duero region and made from Cabernet Sauvignon grapes of the 2017 harvest, was used. The initial wine presented pH 3.89, alcoholic degree 15% (v/v), 5.1 total acidity (g/L tartaric acid), 0.48 volatile acidity (g/L acetic acid), 29 mg/L of free SO2 and 74 mg/L of total SO2, 15.1 colour intensity, % Abs 420 of 38.4, % Abs 520 of 50.1, % Abs 620 of 12.1 and a total polyphenol index of 66 according to OIV, 2003 [33 ]. The colour analysis was measured with UV/Vis Spectrometer Lambda 25 (Perkin Elmer, Singapore). Each measurement was made in duplicate according to Glories [34 (link)].
The wine was barrelled (March of 2018) in 10 barrels (225 L)—4 barrels with High wood-OTR, 4 barrels with Low wood-OTR and 2 control barrels constructed without traceability on the staves, used as control—to age the red wine. The barrels were maintained at 17 °C and with a relative humidity of 70% similar to the common cellar conditions. Wine samples were taken after 6 and 12 months of ageing, after which samples were stored at 18 °C until analysis and the barrels were filled after each sampling. Each sample from each barrel was analyzed in duplicate.
+ Open protocol
+ Expand
10

Characterization of Sym. Radix/AgNPs

Check if the same lab product or an alternative is used in the 5 most similar protocols
The formation of Sym. Radix/AgNPs 1 was checked using a Lambda 25 UV-vis spectrometer (PerkinElmer, Waltham, MA, USA). The analysis was performed in a wavelength range of 200 to 800 nm. Ultrapure water was used as the blank for the UV-Vis experiments.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!