The largest database of trusted experimental protocols

8 protocols using g1313a als

1

Quantification of Analytes by LC-MS

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS analyses were performed on an Agilent 1100 HPLC system equipped with a G13795 degasser, G1312A BinPump, a G1313A ALS and G1316A column oven (COLCOM) (Agilent, Little Island, Cork, Ireland). Separation was obtained on a Kinetex phenyl-hexyl column (2.6 μm, 100 × 2.10 mm) Phenomenex (Macclesfield, Cheshire, United Kingdom). The analytes were eluted under isocratic conditions using amobile phase of 97%water and 3%acetonitrile (both containing 0.1% formic acid). The Agilent single quadrupole MSD settings were as follows: positive electrospray mode, capillary voltage 3500 V, drying gas (N2) 12 L/min at 350 °C, and nebulizer gas (N2) pressure 50 psi. In-source collision-induced dissociation experiments were carried out with an increased fragmentor voltage of 110 V. Samples were dissolved in acetonitrile/water (1:1, containing 0.1% formic acid) at a concentration of 10 μg/mL. The injection volume was 0.5 μL, flow rate was 0.4 mL/min and the column temperature was set at 30 °C. Total run time was 25 min.
+ Open protocol
+ Expand
2

Quantitative LC-MS Analysis of Analytes

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS analyses were performed on an Agilent 1100 HPLC system equipped with a G13795 degasser, G1312A BinPump, a G1313A ALS and G1316A column oven (COLCOM) (Agilent, Little Island, Cork, Ireland). Separation was achieved using a Kinetex phenyl-hexyl column (2.6 μm, 100 x 2.10 mm) from Phenomenex (Macclesfield, Cheshire, United Kingdom). The analytes were eluted under isocratic conditions using a mobile phase of 95% water and 5% acetonitrile (both containing 0.1% formic acid). The Agilent LC-MSD settings were as follows: positive electrospray mode, capillary voltage 3500 V, drying gas (N2) 12 L/min at 350 °C, nebuliser gas (N2) pressure 50 psi, SIM m/z 192, fragmentor voltage 50 V and 150 V. Samples for LC-MS analysis were dissolved in acetonitrile/water (1:1, containing 0.1% formic acid) at a concentration of 10 μg/mL. The injection volume was 0.5 μL, flow rate was 0.2 mL/min and the column temperature was 30 °C. Total run time was 25 min.
+ Open protocol
+ Expand
3

Extraction and Quantification of Soybean Glycosides

Check if the same lab product or an alternative is used in the 5 most similar protocols
Soybeans (GM, 3000 g) were powdered and made circumfluent with 30% ethanol (3 times the weight of the beans) for 3 h at 90 °C. The ethanol supernatants were percolated and lyophilized to yield the final extract (362 g, 12.1%). GE was supplied by Sigma–Aldrich. (Sigma–Aldrich, Saint Louis, MO, USA).
High-performance liquid chromatography (HPLC) was performed using Agilent 1100 series G1379A Degasser, G1312A Bin Pump, G1313A ALS, G1316 column, and G1313A VWD (Agilent Technologies, Santa Clara, CA, USA). As the HPLC column, a Luna (18, 150 × 4.6 mm, 5 μm) column was used, and the wavelength of the ultraviolet absorption detector was 254 nm. As the moving phase, an ammonium sulfate solution (1 → 100) and acetonitrile mixture solution (91.5:8.5, v/v) were used. Thereafter, 10 μL of the sample was injected.
Using this method, the concentration of GE in GM was calculated to be 7.74 µg/g, as described in Figure 6.
+ Open protocol
+ Expand
4

LC-MS Analysis of Organic Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
LC-MS analyses were performed on an Agilent 1100 HPLC system equipped with a G13795 degasser, G1312A BinPump, a G1313A ALS and G1316A column oven (COLCOM) (Agilent, Little Island, Cork). Separation was obtained on an Allure PFP Propyl column (5μm, 50 × 2.1 mm) Restek (Bellefonte, PA, USA). Mobile phase A consisted of 0.1% formic acid in water, whereas, mobile phase B consisted of 0.1% formic acid in acetonitrile. The Agilent LC-MSD settings were as follows: positive electrospray mode, capillary voltage 3500V, drying gas (N2) 12L/min at 350°C, nebulizer gas (N2) pressure 50 psi, scan mode m/z 70-500, fragmentor voltage 50 and 110V. Samples for LC-MS analysis were dissolved in acetonitrile/water (1:1, containing 0.1% formic acid) at a concentration of 10μg/mL. The injection volume was 10.0μL, flow rate was 0.8 mL/min and the column temperature was 30°C. Total run time was 25 min. The following gradient elution program was used: 0-2 min 2% B, followed by an increase to 60% B within 15 min, followed by another increase to 80% B within 18 min before returning to 2% B within 25 min.
+ Open protocol
+ Expand
5

HPLC Analysis of Purine Metabolites

Check if the same lab product or an alternative is used in the 5 most similar protocols
Hypoxanthine, xanthine, 8-aminoxanthine, 8-aminohypoxanthine, 8-aminoinosine, and guanine were measured by HPLC analysis using an Agilent (Santa Clara, CA) HPLC system, which included a HP Agilent Technologies 1100 series HPLC chromatograph equipped with a diode array detector (model G1315B) and autosampler (model G1313A ALS). Samples were heated to 90°C for 1.5 minutes (to inactivate enzymes), vortexed, and centrifuged at 14,000 rpm for 25 minutes at 4°C, and the supernatants were collected and transferred into HPLC vials. Aliquots of samples (2–15 μl) were injected onto a C-18 reverse phase column (Agilent Eclipse Plus C18, 5 μm, 4.6 × 250 mm), which was protected by a guard cartridge. Analysis was conducted in gradient mode [buffer A: 0.15 M KH2PO4 in water (pH 6.0); buffer B: 0.15 M KH2PO4 in 15% acetonitrile (pH 6.0); and linear gradient (percent B): at 0 minutes, 3.0%; from 0 to 9 minutes, 3.0%; from 9 to 25 minutes, to 100.0%; from 25 to 30 minutes, 100.0%; from 30 to 31 minutes, to 3.0%; and from 31 to 40 minutes, 3.0%]. The flow rate was 0.8 ml/min. Purines in the eluate were monitored at 259 nm. Chromatograms were processed and stored in digital form with Agilent OpenLAB CDS software. Standard curves were generated from authentic standards.
+ Open protocol
+ Expand
6

Simultaneous Quantification of Metronidazole and Lidocaine

Check if the same lab product or an alternative is used in the 5 most similar protocols
A reverse-phase HPLC method was developed in-house which allowed the simultaneous determination of the two active drugs in a single assay. This was achieved using an Agilent system comprised of a G1379B Degasser, G1311A QuatPump, G1313A ALS autosampler and G1314A VWD UV detector. An isocratic mobile phase of 80 % ammonium acetate (0.2 mol/L), 20 % acetonitrile plus 0.2 % w/v trifluoroacetic acid was used, along with a Kinetex 5 μm C18 150 x 4.6 mm column (Phenomenex, Macclesfield, UK). The detection wavelength was set at 254 nm, with injection volume 20 µL and flow rate 1.5 mL/min. Under these conditions the retention times of metronidazole and lidocaine were found to be approximately 1.5 and 3.0 min respectively. Calibration curves for metronidazole and lidocaine were constructed using concentrations in the range of 31.25 to 1000 µg/mL in methanol and both agents were fully baseline-resolved. An exemplar chromatogram is shown in Figure 2 and chromatographic parameters were determined as follows: k'metronidazole 6.5, k'lidocaine 14, α 2.15, Rs 4.6.
+ Open protocol
+ Expand
7

HPLC-DAD Analysis of Bilirubin and Biliverdin

Check if the same lab product or an alternative is used in the 5 most similar protocols
HPLC-DAD analyses were carried out using Agilent 1100 Series system (Agilent Technologies, Waldbronn, Germany) fitted with a thermostated autosampler (G1313A ALS) with a 500 μL loop, a quaternary pump (G1311A QUAT PUMP), degasser (G1322A) and with a diode-array detector (G1315A DAD). The evaluation of the collected data was made by ChemStation software (for LC 3D system Rev. B01.03 204, Agilent Technologies 2001-2005). Injection volume for each run was 200 mL. Flow rate was set to 0.6 mL/min, the column temperature was set to 25 °C and the acquisition wavelengths were 377 nm, 407 nm and 457 nm. Bilirubin and biliverdin were separated using a stainless steel BDS Hypersil C18 column (100 mm  4.6 mm I.D.) with a pre-column (4.0 mm  4.6 mm I.D.) with particle size of 2.4 mm and pore size of 120 Å (Thermo Fisher Scientific, Waltham, USA). The isocratic elution was made using 34% of 20 mM NH 4 OAc solution in MeOH.
+ Open protocol
+ Expand
8

Quantification and Characterization of Clonazolam and Nitrazolam

Check if the same lab product or an alternative is used in the 5 most similar protocols
This was used for the quantification of clonazolam in the tablets and to obtain product ions spectra for clonazolam and nitrazolam utilizing in-source CID.
Analyses were performed on an Agilent 1100 HPLC system equipped with a G13795 degasser, G1312A BinPump, a G1313A ALS and G1316A column oven (COLCOM) (Agilent, Little Island, Cork). Separation was obtained on an Allure PFP Propyl column (5 μm, 50 x 2.1 mm) Restek (Bellefonte, PA, USA). Mobile phase A consisted of 0.1% formic acid in water, whereas mobile phase B consisted of 0.1% formic acid in acetonitrile. The flow rate was set at 0.80 mL/min, with an injection volume of 10 µL, and the column temperature was 30°C. The following gradient elution program was used: 0-2 min 2% B, followed by an increase to 60% B within 15 min, followed by another increase to 80% B within 18 min before returning to 2% B within 25 min (total run time was 25 min.). The Agilent LC-MSD mass spectrometer settings were as follows: positive electrospray mode, capillary voltage 3500 V, drying gas (N 2 ) 12 L/min at 350 °C, nebulizer gas (N 2 ) pressure 50 psi, and scan mode m/z 70-500.
To obtain product ions spectra, samples were dissolved in acetonitrile/water (1:1, containing 0.1% formic acid) and the fragmentor voltage was set at 150 V. For the quantification of clonazolam in tablets, the fragmentor voltage was set at 50 V.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!