Before data acquisition and downstream analysis, samples were cut into small pieces, while 1 g of each sample was placed in a 20 mL vial. Then, the samples were placed into a thermoblock at 50 °C for 20 min for headspace generation and 0.5 mL from the headspace was drawn off and injected into the electronic nose. Each sample was transferred to the detector at a constant rate over 120 s. Following that, a cleaning procedure of the detector chamber was carried out until the sensor signals returned to baseline. As sample gases flowed over the sensors, the sensors’ resistance (R) changed. Therefore, a ratio (R−R0)/R0 was used to estimate the changes in sensor resistance, where R0 is the sensor’s resistance baseline and R is the real-time resistance. Twelve maximum response values of each sample from each sensor (
Fox 3000
The FOX 3000 is a laboratory instrument designed for the analysis of volatile organic compounds (VOCs). It features an array of metal oxide sensors that are capable of detecting and identifying a wide range of chemical compounds. The instrument provides quantitative and qualitative data on the detected VOCs, enabling users to analyze and monitor various samples.
Lab products found in correlation
Market Availability & Pricing
Is this product still available?
Get pricing insights and sourcing optionsSpelling variants (same manufacturer)
The spelling variants listed below correspond to different ways the product may be referred to in scientific literature.
These variants have been automatically detected by our extraction engine, which groups similar formulations based on semantic similarity.
Product FAQ
12 protocols using «fox 3000»
Electronic Nose Analysis of Milk Samples
Before data acquisition and downstream analysis, samples were cut into small pieces, while 1 g of each sample was placed in a 20 mL vial. Then, the samples were placed into a thermoblock at 50 °C for 20 min for headspace generation and 0.5 mL from the headspace was drawn off and injected into the electronic nose. Each sample was transferred to the detector at a constant rate over 120 s. Following that, a cleaning procedure of the detector chamber was carried out until the sensor signals returned to baseline. As sample gases flowed over the sensors, the sensors’ resistance (R) changed. Therefore, a ratio (R−R0)/R0 was used to estimate the changes in sensor resistance, where R0 is the sensor’s resistance baseline and R is the real-time resistance. Twelve maximum response values of each sample from each sensor (
Volatile Profile Analysis of Fish Samples
where Rt is the resistance of the sensor at time t and R0 is the baseline resistance (t = 0). The acquisition time was set to 120 s, which was followed by a recovery period of 1080 s so that the sensors returned to the baseline. The maximum sensor resistance was employed for data analysis. Details of the operating conditions of the E-nose can be found elsewhere [30 (link)].
Volatile Compounds Analysis by GC-MS and Electronic Nose
chromatography-mass spectrometry using a modified version of the method
described by Ba et al. (2010) (link).
Approximately 1 g of dry sample (prepared in duplicate) was immediately placed
in a 50 mL headspace vial and heated at 105°C in a drying oven for 10 min
to release the volatile compounds. Prior to extraction, the sample was
calibrated to 60°C in a drying oven for 10 min. The
carboxen®/ polydimethylsiloxane fiber (Supelco,
Sigma-Aldrich) with a diameter of 75 μm was injected into the vial for
extraction for 30 min. Following extraction, the fiber was injected into the
inlet, which was set to 250°C. The split ration of 1:5 was used for
desorbing the volatile compounds for 5 min. Helium was used as the carrier gas
at a flow rate of 1 mL/min. Separation of the individual compound was performed
using a DB5 fused silica column (30 m×0.25 mm inner diameter, 0.25
μm film thickness; J&W Scientific, Folsom, CA, USA) in a gas
chromatograph (7890A, Agilent Technologies). The GC oven was set to operate at
an initial temperature of 40°C for 2 min, increased to 160°C (at
rate of 5°C/min), then to 180°C (at rate of 6°C/min,
holding time of 5 min), and finally to 200°C (at rate of 10°C/min,
holding time of 5 min). The interface and quadruple temperatures were set at
280°C and 150°C, respectively. Volatile compounds were detected
using a mass spectrometer (5975C, Agilent Technologies). The ion source
temperature of the MS was set to 280°C with an electron impact of 70 eV.
A scanning mass range of 50–450 m/z with a scan rate of 1 scan/s was
used. Identification was performed by comparing the experimental spectra with
the National Institute of Standards and Technology (NIST) mass spectral library.
Data are presented as area units (AU)×106 /g.
An electronic nose (FOX3000, Alpha MOS, Toulouse, France) was used for analyzing
the aroma pattern. Dry and heated samples (0.5 g) were placed in a 10 mL
headspace vial and prepared in duplicate. The vial was sealed with a rubber
septa cap (Supelco 29176-U, Sigma-Aldrich). The samples were heated at
60°C for 600 s at an agitation speed of 500 rpm. The 2.5 mL of headspace
gas was extracted with an automatic sampler syringe (HS 100, Alpha MOS) at
65°C, flow-injected into the port of the electronic nose with synthetic
air as the carrier gas (pressure was set to 0.5 bar with 150 mL/min flow rate)
and detected by a metal oxide sensor array system with an acquisition time of
150 s. The following sensors were chosen (T30/1, P10/1, P10/2, P40/1, T70/2,
PA2) as the sensitivity against fat-derived volatile compounds are high. The
sensor resistance ratio (r−r0)/r0 was calculated (r
is the real-time resistance and r0 is the
sensor’s resistance baseline). The time taken to return
to baseline after acquisition was 1,080 s. The maximum resistance ratio was
considered as the data value of a single measurement.
Headspace E-nose Analysis of Enzymatic Hydrolysates
Flavor Analysis of Dried Okra
About PubCompare
Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.
We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.
However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.
Ready to get started?
Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required
Revolutionizing how scientists
search and build protocols!