The largest database of trusted experimental protocols

Ics 3000

Manufactured by Thermo Fisher Scientific
Sourced in United States

The ICS-3000 is an ion chromatography system designed for the separation and analysis of ionic species. It provides precise and reliable ion detection and quantification for a variety of applications.

Automatically generated - may contain errors

257 protocols using ics 3000

1

Fecal Organic Acid Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The fecal samples were thawed at 4°C and mixed homogeneously. Approximately 0.5 g of the sample was placed in a 10 ml polypropylene tube and 8 ml of ultrapure water was added, then centrifuged at 12,000 g for 10 min at 4°C after sonication for 30 min in an ice-water bath (mixing per 10 min). The supernatant was extracted and diluted 50 times with ultrapure water, filtered through a 0.22 mm membrane, and added to the injection vial. The analysis was carried out using ICS-3000 ion chromatography (ICS-3000, Thermo Scientific, USA), the external standard solution containing eight organic acids was obtained from Sigma-Aldrich (Saint Louis, USA). A variety of organic acids were separated by an AS11 analytical column (250 × 4 mm) and an AG11 guard column (50 × 2 mm) under the mobile phase elution conditions: potassium hydroxide gradient, 0–5 min, 0.8–1.5 mM; 5–10 min, 1.5–2.5 mM; 10–15 min, 2.5 mM, 1 ml/min flow rate.
+ Open protocol
+ Expand
2

Ion Chromatography Analysis of Soluble Compounds

Check if the same lab product or an alternative is used in the 5 most similar protocols
Samples were melted in a class 100 clean room at LGGE-CNRS laboratory (Grenoble, France). They were then transferred into Dionex glass vials previously rinsed with ultra-pure Millipore water (conductivity > 18.2 mΩ, TOC < 10 ng/g) and analyzed less than 24 h after melting. Analyses were performed by conductivity-suppressed ion chromatography using a Dionex ICS 3000© apparatus and a Dionex AS40© autosampler placed in the clean room facilities. Different chemical parameters were measured during this study (e.g., major/minor ions, organic acids, and pH). Soluble anions (methyl sulfonic acid (MSA), SO4, NO3, Cl) and cations (Na, NH4, K, Mg, Ca) and organic acids were analyzed by ionic chromatography (IC, Dionex ICS3000). AS/AG 11HC and CS/CG 12A columns were used for anions and cations analyses, respectively. All chemical analyses were carried out at on the airOsol platform of the IGE laboratory in Grenoble, France. This data set can be found in Supplementary Table S1. The following parameters were used for statistical analyses [Organic acids (oxalate, lactate, glutarate, propionate, succinate, formate, acetate), NO3, NH4+, SO42–, mercury, fluoride, calcium, magnesium, bromide, strontium, lithium, sodium, chloride, potassium, number of particles, methyl sulfonic acid (MSA)] and pH. For values below the detection limit, we used the detection limit divided by 2.
+ Open protocol
+ Expand
3

Polysaccharide Hydrolysis and Monosaccharide Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The polysaccharide sample (10 mg) was hydrolyzed with 4 mol/L trifluoroacetic acid at 120 °C in a sealed tube for 4 h, and the hydrolysate was dried with nitrogen to remove excess trifluoroacetic acid. The polysaccharide hydrolysate was dissolved with 10 mL distilled water and then passed through a 0.2 μm microporous filter. Chromatographic separation conditions of fucose (Fuc), rhamnose (Rha), arabinose (Ara), galactose (Gal), glucose (Glc), fructose (Fru), glucuronic acid (GlcA) and galacturonic acid (Gala): Ion Chromatography (IC, DIONEX ICS-3000, California, USA), CarboPacTMPA10 (4 × 250 mm) analytical column, using sodium acetate (1 mol/L), sodium hydroxide (200 mmol/L) and water as the mobile phase at a flow rate of 1 mL/min. Chromatographic separation conditions of xylose (Xyl) and mannose (Man): IC (DIONEX ICS-3000, California, USA), CarboPacTMPA20 (3 × 150 mm) analytical column, using sodium hydroxide (250 mmol/L) and water as the mobile phase at a flow rate of 0.5 mL/min. The column temperature was 35 °C, and a 10 μL sample was injected.
+ Open protocol
+ Expand
4

Characterization of Galacto-oligosaccharides

Check if the same lab product or an alternative is used in the 5 most similar protocols
We characterized GOS-60.6, GOS-63.5, GOS-71, and GOS-97. The amount of Glc, Gal, Lac, and transgalactosylated oligosaccharides in the samples was studied by HPAEC-PAD analysis. The profiles of the oligosaccharide products were analyzed on a Dionex ICS-3000 work station (Dionex, Amsterdam, The Netherlands) equipped with an ICS-3000 pulse amperometric detection (PAD) system and a CarboPac PA-1 column (250 × 2 mm; Dionex), using a gradient adapted from van Leeuwen et al. 14
+ Open protocol
+ Expand
5

Sweat Ion Concentration Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
All sample analysis with the HPLC took place in a biochemistry laboratory at room temperature. CENTRIFUGE sweat samples were immediately prepared and placed in the HPLC for analysis. After syringe extraction, the SYRINGE sweat samples sat at room temperature in the biochemistry laboratory for at least 30 min prior to analysis with HPLC. Sweat samples were then diluted 1:100 with 18 MΩ ultrapure water (Milli‐Q, Millipore, Billerica) to a total volume of 1 mL. The ion chromatography system consisted of the following Dionex equipment: ICS‐3000, Ion Pac CS12A column, CR‐TC, Ion Pac CG12A guard, CSRS 300 suppressor and conductivity detector (Thermo‐Fisher, Waltham). Methodology included a 25 μL injection at a 1.0 mL/min flow rate for 15 min, 59 mA suppression, and 20 mmol/L methanesulfonic acid isocratic eluent. Four‐point calibration standards were prepared using Dionex combined 6‐cation standard stock solution. The calibration range of [Na+] and [K+] were 4–40 ppm and 0.5–5 ppm (i.e., 400–4000 ppm and 50–500 ppm undiluted), respectively. Validation samples were run daily and periodically throughout a given sequence.
+ Open protocol
+ Expand
6

Volatile Fatty Acid Analysis in Piglet Cecum

Check if the same lab product or an alternative is used in the 5 most similar protocols
The cecal contents of piglets were thawed at 4 °C and mixed, approximately 0.5 g of the sample was weighed into a 10-mL centrifuge tube, adding 8 mL of ultrapure water, sonicated in an ice water bath for 30 min (mixing every 10 min), then centrifuged at 15,000 × g for 10 min. The supernatant was diluted 50 times with ultrapure water, filtered through a 0.22 mm membrane, transferred to the 2-mL injection vial and analyzed the volatile fatty acid using a high-performance ion chromatography analyzer (ICS-3000, Thermo Scientific, USA).
+ Open protocol
+ Expand
7

Quantitative Colistin Analysis by Ion Chromatography

Check if the same lab product or an alternative is used in the 5 most similar protocols
Chromatographic analysis was performed using a Dionex ICS-3000 ion chromatography system (Thermo Scientific, Gloucester, UK) equipped with a Dionex RF-2000 fluorescence detector and a Dionex AS autosampler. Data was collected and processed using Chromeleon 6.8 software. Separation was achieved on an XSelect CSH C18 column (130 Å, 3.5 µm, 3.0 × 150 mm) connected to an XSelect guard column (130 Å, 3.5 µm, 2.1 × 5 mm) held at 30 °C with a flow rate of 0.8 ml/min (Waters; Wilmslow, UK). Samples were kept in the autosampler at 4 °C and eluted (30 µl) using a mixture of ACN, THF and MQ H2O (82:2:16 v/v/v) as a mobile phase over 20 min. The eluted peaks were recorded at excitation and emission wavelengths of 260 nm and 315 nm, respectively, with the gain set at 16 and response at 0.5. Colistin concentration in the lower Transwell compartment after diffusion through AS medium for 48 h was calculated as follows and expressed as mean ± SD (n = 3): Colistinμg/ml=1/CrA/ISIA/IISQIS where A = analyte, IS = internal standard, Cr = slope of the calibration curve, I = area under the curve and Q = quantity of IS in the sample.
A colistin calibration curve was generated using a linear regression method, by plotting the ratio of colistin (B and A) to IS (B2 and B1) using their summed areas under the curve versus the concentration ratio of colistin to IS (µg/ml).
+ Open protocol
+ Expand
8

Wheat Grain Soluble Sugar and Starch Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
Wheat grains were milled, and 25 mg of each sample were added to 0.5 mL of 100% ethanol then another 0.5 mL of 80% ethanol was added and heated in a thermomixer (70 °C, 90 min, 1100 rpm). The mixture was centrifuged (22 °C, 10 min, 20,800× g) and the supernatant was used for the determination of soluble sugars (glucose, fructose and sucrose) concentration, using an ionic chromatographer (ICS-3000, Thermo ScientificTM, Waltham, MA, USA). Reference was made to sugar standards of known concentrations (50 mM). The pellet was used to determine starch content. Starch was solubilized by adding potassium hydroxide (KOH) (0.2 N) to the pellet, and the pH was adjusted to 4.8 with acetic acid (0.1 N). Quantification was performed with the kit containing the enzyme amyloglucosidase (R-Biopharm, AG; Darmstadt, Germany) measuring the absorbance at 340 nm with a spectrophotometer.
+ Open protocol
+ Expand
9

Oligosaccharide Separation and Analysis

Check if the same lab product or an alternative is used in the 5 most similar protocols
The TVE extract that contained the oligosaccharide mixture was ultrafiltered on a 3kDa cut-off membrane to separate the very low molecular weight fraction (e.g., glucose and xylose). The permeate containing monosaccharide was analyzed by a high pressure ion chromatography system (ICS3000, Thermo Fisher Scientific, Milan, Italy) that was equipped with a pulsed amperometric detector (reference electrode Ag–AgCl; measuring electrode Au) and anion exchange column (Carbopac PA1, Thermo Fisher Scientific, Milan, Italy). Isocratic conditions (NaOH 16mM) were used, 25 µL of sample were injected at a flow rate of 0.9 mL/min and a temperature of 25 °C.
+ Open protocol
+ Expand
10

Carbohydrate Analysis by Bio-LC

Check if the same lab product or an alternative is used in the 5 most similar protocols
Bio-LC system (ICS-3000; Thermo-Fisher, USA) equipped with a CarboPac PA1 column (4 × 250 mm) and an electrochemical detector (ED40) was utilized for the analysis of enzymatic hydrolysates. The samples were eluted with a linear gradient from 100% 150 mM NaOH (buffer A) to 15% buffer B (600 mM sodium acetate in buffer A) over 40 min. The flow rate of the mobile phase maintained at 1.0 ml/min through the analysis.
+ Open protocol
+ Expand

About PubCompare

Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.

We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.

However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.

Ready to get started?

Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required

Sign up now

Revolutionizing how scientists
search and build protocols!